magnification: (a) objective lens 40X (Cb) adjustable magnifier 1.5X (c) eye piece 10X (d) total system 600x° (base of viewing head fitted “with adjustable magnification changer unit and a focusing bertrand lens) illumination system: low voltage 6V1i8W illuminator with field centrable field iris type of graticule used: British standard graticule, defined area at plane of focus is 75 x 53 micrometers diameter of full field of view: 263 micrometers REAGENTS Mounting Solution: Triacetin which is free of particles and colour is required. Filter materials identical to that used in eenpiine is also required.” MOUNTING PROCEDURES Sample Mounting: Using the scalpel the exposed filter sample is carefully sliced into the appropriate number of sectors ase required for analysis of the sample. With a glass rod supplied with the Wheaton Balsam bottle, apply a small drop of filtered mounting solution (Triacetin) to a cleaned glass slide or the mounting solution may be dispensed through a solvent resistant membrane filter held in a syringe filter holder. Filtration of the cleaning agent is particularly important if asbestos contamination is possible, and if relatively . low dust levels are to be meeaned. It is important to avoid using an : excessive amount of solution so as to avoid migration of the asbestos particles. With another clean glass rod, or use the tip of a syringe to spread spot into circle, spread the spot. of mounting solution so that it coincides